JASZCZAK STANDARD SPECT PHANTOM ™
ECT phantoms with protruding flanged top and 3.2 mm cylinder wall thickness.
PRODUCT DESCRIPTION
ECT phantoms with protruding flanged top and 3.2 mm cylinder wall thickness.
MAIN APPLICATIONS
-
For use with SPECT and PET systems
-
System performance evaluation (collimator, artifacts, calibration, reconstruction parameters)
-
Acceptance testing
-
Routine quality, assurance and control
-
Evaluation of center-of-rotation error
-
Evaluation of non-uniformity artifact
-
Evaluation of changes of radius-of-rotation on spatial resolution
-
Evaluation of reconstruction filters on spatial resolution
-
Evaluation of attenuation and scatter compensation
-
Research
SPECIFICATIONS OF CYLINDER
-
All clear material: PMMA
-
Cylinder inside diameter: 21.6 cm
-
Cylinder inside height: 18.6 cm
-
Cylinder wall thickness: 3.2 mm
SPECIFICATIONS OF INSERT
-
All clear material: PMMA
-
Rod diameters: Vary with insert
-
Height of rods: 8.8 cm
-
Solid sphere diameters: Vary with inert
-
Height of center of spheres from base plate: 12.7 cm